Mostrar el registro sencillo del ítem

dc.contributor.authorGordillo, Rafael M.
dc.contributor.authorGutiérrez-Aroca, Juan-Bautista
dc.contributor.authorCasal Román, Manuel
dc.date.accessioned2018-12-19T13:22:57Z
dc.date.available2018-12-19T13:22:57Z
dc.date.issued2005
dc.identifier.urihttp://hdl.handle.net/10396/17599
dc.description.abstractThe purpose of this study is to evaluate the usefulness of the new real-time PCR COBAS TaqMan 48 analyzer, comparing it to the existing COBAS AMPLICOR HBV MONITOR based on conventional PCR technology. The study used 104 samples from different patients. No differences were found in the sensitivity of the tests. There was an excellent correlation between the sample with a viral load within the dynamic range of the two tests (r = 0.938). The COBAS TaqMan test has a wider linear range, and this fact enables quantifying of the viral load without diluting the sample.es_ES
dc.format.mimetypeapplication/pdfes_ES
dc.language.isoenges_ES
dc.publisherAmerican Society for Microbilogyes_ES
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/4.0/es_ES
dc.sourceJournal of Clinical Microbiology 43 (7), 3504-3507 (2005)es_ES
dc.subjectPCR COBASes_ES
dc.subjectVirologyes_ES
dc.titleEvaluation of the COBAS TaqMan 48 Real-Time PCR System for Quantitation of Hepatitis B Virus DNAes_ES
dc.typeinfo:eu-repo/semantics/articlees_ES
dc.relation.publisherversionhttps://doi.org/10.1128/JCM.43.7.3504-3507.2005es_ES
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses_ES